Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
Introduction: Corneal curvature (CC) is an important anterior segment parameter. This study compared CC measurements conducted with two optical devices in pterygium eyes.
Methods: Sixty pterygium eyes of 30 patients were enrolled in this study. CC was measured three times with the optical biometer and topography-keratometer Tomey OA-2000 (Tomey Corporation, Nagoya, Japan), then with partial optical coherence interferometry (PCI) IOL Master 500 (Carl Zeiss Meditec, Dublin, CA) and data were statistically analysed.
Results: The measurements revealed in a mean CC of 43.86 ± 1.57 D with Tomey OA-2000 and 43.84 ± 1.55 D with IOL Master. Distribution of data is normal, and no significance difference in CC values was detected (P = 0.952) between the two devices. Correlation between CC measurements was highly significant (r = 0. 99; P < 0.0001). The mean difference of CC values between devices was 0.017 D and 95% limit of agreement was -0.088 to 0.12. Duration taken for measurements with the standard biometer IOL Master was longer (55.17 ± 2.24 seconds) than with Tomey OA-2000 (39.88 ± 2.38 seconds) in automatic mode. Duration of manual measurement with Tomey OA-2000 in manual mode was shorter (28.57 ± 2.71 seconds).
Conclusion: In pterygium eyes, CC measured with Tomey OA-2000 and IOL Master showed similar values, and high correlation was observed between these two devices. This shows that both devices can be used interchangeably. Tomey OA-2000 is better in terms of faster to operate and has its own topography systems.
Keywords: Pterygium; Corneal Curvature; IOL Master 500; Tomey OA-2000; Corneal Topography
- Altan-Yaycioglu, R. et al., 2013. Astigmatic changes following pterygium removal: comparison of 5 different methods, Indian J Ophthalmol, 61(3), pp.104–108.
- Bandyopadhyay, R. et al., 2010. Ocular surface disorder in pterygium; Role of conjunctival immersion cytology, Indian J Pathol Microbiol, 53(4), pp.692–695.
- Barkana, Y. et al., 2005. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry, J Cataract Refract Surg, 31(9), pp. 1729–1735.
- Bueno-Gimeno, I. et al., 2013 Anterior chamber depth measurement in teenagers. Comparison of two techniques, J Optom, 6(3), pp.161–166.
- Chen, Y.A., Hirnschall, N. & Findl, O. 2011. Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer, J Cataract Refract Surg, 37(3), pp.513-517.
- Dehnavi, Z. et al., 2015. Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOLMaster, and Javal Keratometer, Middle East Afr J Ophthalmol, 22(2), pp.233–237.
- Elbaz, U. et al., 2007. Comparison of different techniques of anterior chamber depth and keratometric measurements, Am J Ophthalmol, 143(1), pp.48–53.
- Hashemi, H. & Mehravaran, S. 2007. Corneal changes after laser refractive surgery for myopia: Comparison of Orbscan II and Pentacam findings, J Cataract Refract Surg, 33(5), pp.841–847.
- Hashemi, H. & Mehravaran, S. 2010. Day to day clinically relevant corneal elevation, thickness, and curvature parameters using the orbscan II scanning slit topographer and the pentacam scheimpflug imaging device, Middle East Afr J Ophthalmol, 17(1), pp.44–55.
- Holzer, M.P., Mamusa, M. & Auffarth, G.U. 2009. Accuracy of a new partial coherence interferometry analyser for biometric measurements, Br J Ophthalmol, 93(6), pp.807–810.
- Huang, J. et al., 2017. Repeatability and interobserver reproducibility of a new optical biometer based on swept-source optical coherence tomography and comparison with IOLMaster, Br J Ophthalmol, 101(4), pp.493-498.
- Huynh, S.C. et al. 2006 An evaluation of keratometry in 6-year old children, Cornea, 25, pp.383–387.
- Khairat, Y.M. et al., 2013. Evaluation of corneal changes after myopic LASIK using the Pentacam®, Clin Ophthalmol, 7, pp.1771–1776.
- Kheirkhah, A. et al. 2012. Effects of pterygium surgery on front and back corneal surfaces and anterior segment parameters, Can J Ophthalmol, 47(5), pp.423–428.
- Kheirkhah, A. et al., 2012. Effects of pterygium surgery on front and back corneal astigmatism, Int Ophthalmol, 32, pp.251–257.
- Kim, E.J. et al., 2015. Repeatability of posterior and total corneal curvature measurements with dual Scheimpflug-Placido tomographer, J Refract Surg, 41(12), pp.2731-2738.
- McAlinden, C. & Moore, J.E. 2010. Comparison of higher order aberrations after LASIK and LASEK for myopia, J Refract Surg, 26(1), pp.45–51.
- McAlinden, C. 2012. Corneal refractive surgery: past to present, Clin Exp Optom, 95(4), pp.386–398.
- Tennen, D.G., Keates, R.H. & Montoya, C. 1995. Comparison of three keratometry instruments, J Cataract Refract Surg, 21, pp.407–408.
- Mehravaran, S. et al., 2014. Keratometry with five different techniques: a study of device repeatability and inter-device agreement, Int Ophthalmol, 34(4), pp.869–875.
- Nakakura, S. et al., 2012. Comparison of anterior chamber depth measurements by 3-dimensional optical coherence tomography, partial coherence interferometry biometry, Scheimpflug rotating camera imaging, and ultrasound biomicroscopy, J Cataract Refract Surg, 38(7), pp.1207–1213.
- Norrby, S. 2008. Sources of error in intraocular lens power calculation, J Cataract Refract Surg, 34(3), pp.368–376.
- Ozsutcu, M. et al., 2014. Tear osmolarity and tear film parameters in patients with unilateral pterygium, Cornea, 33(11), pp.1174–1178.
- Rosa, N. et al., 2006. Anterior chamber depth measurement before and after photorefractive keratectomy: comparison between IOL master and Orbscan II, Ophthalmology, 113(6), pp.962–969.
- Rosa, N. et al., 2011. Reliability of the IOLMaster in Measuring Corneal Power Changes After Hyperopic Photorefractive Keratectomy, J Cataract Refract Surg, 27(4), pp.293–298.
- Salouti, R. et al., 2010. Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II, Optometry, 81(1), pp.35–39.
- Savini, G. et al., 2016. Estimating the Preoperative Corneal Power With Scheimpflug Imaging in Eyes That Have Undergone Myopic LASIK, J Refract Surg, 32(5), pp.332-336.
- Sheppard, A.L. & Davies, L.N. 2010. Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500, Ophthal Physiol Opt, 30(2), pp.143–151.
- Shirayama, M. et al., 2009. Comparison of corneal powers obtained from 4 different devices, Am J Ophthal, 148(4), pp.528–535.
- Twa, M.D., Coral-Ghanem, C. & Barth, B. (2003) ‘Corneal topography and contact lenses’ in Mannis, M.J., et al., (eds.) Contact lenses in ophthalmic practice. New York: Springer-Verlag; 2003.
- Vives, P.P. et al., 2013. Topographic corneal changes in astigmatism due to pterygium’s limbal-conjunctival autograft surgery, J Emmetropia, 4, pp.13–18.
- Wang, Q. et al., 2017. Central corneal thickness and its relationship to ocular parameters in young adult myopic eyes, Clin Exp Optom, 100(3), pp.250-254.
- Wang, W. et al., 2017. Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography, Sci Rep, 7(1), pp.13736. doi: 10.1038/s41598-017-13800-7.
- Whang, W-J., Byun, Y-S. & Joo, C-K. 2012. Comparison of refractive outcomes using five devices for the assessments of preoperative corneal power, Clin Exp Ophthalmol, 40, pp.425–432
This work is licensed under a Creative Commons Attribution 4.0 International License.